a50caa263718426e8498e762bb455865-560×500

UPC5200/UPC5210 High Pressure Calibration Standard

  • Available in enclosure with carrying handles (UPC5200) or 19 inch wide rack-mountable version (UPC5210)
  • Two sets of ranges to choose from: 10,000/5,000/2,000 psi and 5,000/2,500/1000 psi
  • Three independent selectable ranges in each instrument
  • ±0.05% F.S. accuracy of each independent selectable range
  • Available with Light Emitting Diode (LED) or Liquid Crystal Display (LCD)
  • Pressure display resolution 0.02% F.S. typical for each range
  • Different technicians can calibrate the same Device-Under-Test and get identical results
  • Electronically controlled 10:1 intensifier system
  • Easy to operate, training can be accomplished in 10 minutes
  • Easy to carry; approximately 51 lb
  • Adjustable overpressure protection above input pressure
  • Precision pressure vernier control
  • Voltage/current display mode for transducer testing
  • 32 VDC excitation output
  • Non-destructible, finger tip control micrometering valves for pressure control
  • Automatic pushbutton calibration, microprocessor controlled and maintained
  • Quick-disconnect fitting and hose included
  • Pneumatic only; no hydraulics or distilled water messes
  • External portable pressure cylinder included with UPC5200
  • Calibration certificate included; traceable to the National Institute of Standards and Technology(NIST)
Cotizar por whatsapp

Descripción

Available Calibration Modes:
Gauge

Overall accuracy:
±0.05% F.S. maximum. Accuracy statement includes all effects of linearity, hysteresis, repeatability and ambient temperature.

Power:
115 or 230 VAC 50/60 Hz

Temperature Range:
Operating: 40°F to 140°F (4°C to 60°C)
Storage: 0°F to 185°F (-17°C to 85°C)
Relative humidity: 95%, non-condensing

Pressure Media:
Clean dry gaseous nitrogen, standard

External Pressure Cylinder Supplied:
Capacity: 88 standard cubic feet
Volume: 960 cubic inches
Rating: 2216 psi

Regulated Pressure Gauge:
Size: Two and one half-inch diameter
Range: 0-1500

Pressure Media Filter:
Rating: 7 microns, nominal
Type: Field replaceable filter cartridge contained within a stainless steel outer case

Control Valves:
Type: Micro-metering with positive shut off, adjustable stop and replaceable seat, non-destructible
Material: 300 series stainless steel outer body

Internal Piping:
Tubing: 1/4 in. O.D. and 1/8 in. O.D., 300 series stainless steel, seamless
Couplings: 300 series stainless steel Swagelok®

Output Test Port:
Style: Quick-disconnect type
Pressure Rating: 10,000 psig minimum, connected or disconnected

Pressure Hoses:
Quantity supplied: Two; one input and one output
Length: 4 feet, nominal

Vernier Control:
Vernier Volume: 0.4 cubic inch, nominal
Vernier Mechanical Rotation: 15 turns, nominal
Adjustable Sensitivity: 0.005 psi, minimum

Booster Pump Solenoid Valve:
Type: Three-way, normally closed, electronically controlled
Pressure Rating: 1500 psig
Power Requirements: 115 or 230 VAC, 50/60 Hz, 10 watts

Accumulator Gauge:
Size: Two and one half-inch diameter
Range: 0-15,000 psi

Input Pressure Regulator:
Pressure Rating: 3000 psig maximum
Output Pressure Range: 0-1000 psi
Construction: Self-relieving, single stage regulator

UPC5200 Internal Pressure Sensor:
Type: Bonded, metal foil strain gauge or equal
Sensitivity: 3 mV/V, nominal
Construction: Completely weld sealed stainless steel outer body and pressure cavity

External Carrying Case:
Type: Deep-drawn aluminum case with cover and carrying handles
Material Thickness: 0.090 in. nominal
Finish: Grey polyurethane baked-enamel paint, textured finish

Current Display:
0-20.000 mA counting by 0.005 mA

Weight:
UPC5200: 51 lb (23.1 kg) including all hoses and cables
External cylinder 5 lb (2.26 kg)