UPC5100/UPC5110 Pressure Vacuum Calibration Standard
- Available in self-contained enclosure with carrying handles (UPC5100) or 19 inch wide rack-mountable version (UPC5110)
- Five independent selectable ranges in each instrument ±0-28 inHg, ±0-387 inH2O, 0-2771 inH2O, ±0-96 kPa, 0-100 psi
- Available with Light Emitting Diode (LED) or Liquid Crystal Display (LCD)
- Pressure/vacuum display resolution 0.02% F.S. typical for each range
- ±0.05% F.S. accuracy for each range
- State-of-the-art push button calibration—no analog potentiometers or adjustments required
- Different technicians can calibrate the same Device-Under-Test and get identical results
- Easy to operate, training can be accomplished in 10 minutes
- Lightweight – less than 35 pounds
- Adjustable overpressure protection
- Precision vernier control
- Self-contained nitrogen cylinder and vacuum converter provides 27 inHg (w/out external pump) to 100 psi
- Voltage/current display mode for transducer testing
- Quick-disconnect fitting and hose included
- Pneumatic only; no hydraulic fluid or distilled water messes
- Calibration certificate included; traceable to the National Institute of Standards and Technology (NIST)
- Descripción
Descripción
Available Calibration Modes:
Gauge and vacuum
Overall accuracy:
±0.05% F.S. maximum. Accuracy statement includes all effects of linearity, hysteresis, repeatability and ambient temperature
Temperature Range:
Operating: 40°F to 140°F (4°C to 60°C)
Storage: 0°F to 185°F (-17°C to 85°C)
Relative humidity: 95% non-condensing
Pressure Media:
Clean dry gaseous nitrogen or clean dry air
Pressure Media Filter:
Rating: 7 microns nominal
Type: Field replaceable filter cartridge
Control Valves:
Type: Micro-metering with positive shut off, adjustable stop and replaceable seat, non-destructible
Material: Aluminum/stainless steel
Internal piping:
Tubing: 5/32 in. nylon tubing, 1/8 in. copper
Fittings: Brass
Test Port:
Style: Quick-disconnect type
Pressure rating: 600 psi maximum, connected or disconnected
Pressure Hose:
Quantity supplied: One output
Length: 5 feet, nominal
Vernier Control:
Volume: 0.25 cubic inches
Vernier Mechanical Rotation: 16 turns, nominal
Adjustable Sensitivity: 0.0005 psi, minimum
Material: Aluminum/stainless steel
Pressure Limit Monitor Gauge:
Size: Two-inch diameter
Range: 160 psi
Pressure supply gauge: 3000 psi
Regulator: 100 psi max
Construction: Self-relieving, single stage regulator
Internal Pressure Sensor:
Type: Bonded, metal foil strain gauge or equal
Sensitivity: 3 mV/V, nominal
Construction: Completely weld sealed stainless steel outer body and pressure cavity
UPC5100 External Carrying Case:
Type: Aluminum case with cover and carrying handle
Material thickness: 0.090 in. nominal
Finish: Grey polyurethane baked-enamel paint, textured finish
Current Display:
0-20.000 mA counting by 0.005 mA
Vacuum Converter:
Capable of generating 27 inches Hg vacuum stand alone or 30 inches Hg with external vacuum pump
Weight:
UPC5100: Less than 35 lb (15.9 kg) including all hoses and cables