a4d9aabbb26b4871a6e74093d8e712ed-560×500

UPC5000/UPC5010 Pressure Calibration Standard

  • Available in self-contained enclosure with carrying handles (UPC5000) or 19 inch wide rack-mountable version (UPC5010)
  • Six sets of ranges to choose from: 2,000/1,000/400 psi, 1,000/500/200 psi, 500/250/100 psi, 100/50/20 psi, 50/25/10 psi and 15/7.5/3 psi
  • Three independent selectable ranges in each instrument
  • ±0.05% F.S. accuracy of each independent selectable range
  • Available with Light Emitting Diode (LED) or Liquid Crystal Display (LCD)
  • Pressure display resolution 0.02% F.S. typical for each range
  • State-of-the-art push button calibration – No analog potentiometers or adjustments required
  • Lightweight, less than 33 pounds
  • Adjustable overpressure protection
  • Precision pressure vernier control
  • Self-contained 2,215 psi nitrogen cylinder
  • Voltage/current display mode for transducer testing
  • Supplies DC loop power
  • Non-destructible, finger tip control micrometering valves for pressure control
  • Quick-disconnect fitting and hose included
  • Different technicians can calibrate the same Device-Under-Test and get identical results
  • Calibration Certificate included; traceable to the National Institute of Standards and Technology (NIST)
  • Pneumatic only; no hydraulic fluid or distilled water messes
Cotizar por whatsapp

Descripción

Available Calibration Modes:
Gauge, absolute or gauge and absolute

Overall accuracy:
±0.05% F.S. maximum. Accuracy statement includes all effects of linearity, hysteresis, repeatability and ambient temperature.

Power:
115 or 230 VAC 50/60 Hz and 12 volt, rechargeable internal, sealed lead acid battery, standard

Temperature Range:
Operating: 40°F to 140°F (4°C to 60°C)
Storage: 0°F to 185°F (-17°C to 85°C)
Relative humidity: 95%, non-condensing

Pressure Media:
Clean dry gaseous nitrogen, standard

Internal pressure cylinder:
Capacity: 7.0 standard cubic feet
Volume: 80 cubic inches
Rating: 2215 psi

Pressure supply gauge: Size: Two-inch diameter
Range: 0-3000 psig

Overpressure Rupture Disc:
Rating: 3000 psig, nominal
Material: Stainless steel outer case

Pressure Media Filter:
Rating: 7 microns, nominal
Type: Field replaceable filter cartridge

Control Valves:
Type: Micro-metering with positive shut off, adjustable stop and replaceable seat, non-destructible
Material: Aluminum/stainless steel

Internal piping:
Tubing: 1/8 in. O.D., 0.028 in. wall thickness, copper, seamless
Couplings: Brass Swagelok®

Fill and Test Ports:
Style: Quick-disconnect type
Fill Port Pressure Rating: 3000 psig connected, 2000
psig disconnected
Test Port Pressure Rating: 5000 psig

Pressure Hoses:
Quantity supplied: Two; one input and one output
Length: 5 feet, nominal

Vernier Control:
Volume: 0.25 cubic inches
Vernier Mechanical Rotation: 16 turns, nominal
Adjustable Sensitivity: 0.0005 psi, minimum
Material: Aluminum/stainless steel

Pressure Limit Monitor Gauge:
Size: Two-inch diameter
Range: As appropriate for UPC5000/UPC5010 model range utilized

Pressure Limit Control:
Pressure Rating: 3000 psig maximum (inlet)
Output Pressure Range: As appropriate for UPC5000/UPC5010 model range utilized
Construction: Self-relieving, single stage regulator

Internal Pressure Sensor:
Type: Bonded, metal foil strain gauge or equal
Sensitivity: 3 mV/V, nominal
Construction: Completely weld sealed stainless steel outer body and pressure cavity

UPC5000 External Carrying Case:
Type: Aluminum case with cover and carrying handle
Material thickness: 0.090 in. nominal
Finish: Grey polyurethane baked-enamel paint, textured finish

Current Display:
0-20.000 mA counting by 0.005 mA

Weight:
UPC5000: Less than 33 lb (14.9 kg) including all hoses and cables